>  >  > Heating
 

Kleinfeld
Setaflash-Tester, Series 3, with Peltier Cooling

Setaflash-Tester, series 3, semi-automatic, closed cup, rapid equilibrium test mode.

The Setaflash-Tester uses an gas ignitor and test cup with lid and shutter. All functions on the series 3 flash point tester are accessed via a two button keypad and the Seta Multifunctional and Rotational Test (SMART) control feature. The digital display shows test parameters, instrument status and the test result.

Application:
Determination of the flash-point and classification of mineral oil, bio-diesel, colour, glue, resolvent or other flammable sustances.

Standards:
ASTM D1655, ASTM D3278, ASTM D3828, ASTM E502, BS 3900 Part A13, BS 6664 Parts 3 & 4, BS EN 456, CHIPS Regulations, DEF STAN 91-91, EPA 1020 A & B, IP 303 (obs), IP 523, IP 524, ISO 3679, ISO 3680, UN Class 3 Non-viscous Flammable Liquids.
Sample size:2 ml for flash points up to 100 °C
4 ml for flash points above 100 °C
Test duration:1 min below 100 °C
2 min above 100 °C
Dimensions (W x D x H):280 x 260 x 260 mm
Others gas:max. 3 kPa
Power:200 W
Mains supply:single phase, 110 / 120 V or 220 / 240 V switchable, 50 / 60 Hz
There are 2 modes of flash point operation: Flash / No Flash and ramp. The ramp mode increases the sample temperature at a rate of 2 °C / min. Barometric pressure is measured by an internal barometer and the correction is automatically applied to the test result. Peltier cell technology is used for flashpoint tests above ambient and for heating and cooling, ensuring precise temperature control and rapid cooldown after a test is completed. Automatic flash detection.
Technical Data:
Temperature range:10 to 130 °C
Weight:5 kg
Test mode:rapid equilibrium and ramp
Ramp rate:2 °C / min
Barometer correction:automatic
Cooling:Peltier cooling
 TypeDescriptionPUOrder No.in Stock 
 Setaflash-Tester with Peltier coolingrapid equilibrium and ramp mode15436598-on demand
Accessories are available on request.