Incubators INplus / IFplus

Models IN / INplus with natural convection.
Models IF / IFplus with forced air circulation.


Incubators for applications in research, medicine, pharmaceutics and food technology. Organic chamber loads require gentle heating. For this reason, the heating and control system are especially optimised for low temperatures of up to + 80 °C. To prevent temperature overshoots, temperature is increased within a very narrow control range and kept exactly at the setpoint value.
  • With all-round surface heating and temperature control system.
  • Fresh air is preheated.
  • Acc. to DIN 12880:2007-05, EN 61010-1 (IEC 61010-1), 61010-2-010.

AtmoCONTROL - control and logging software
For direct setting parameters as well as process time. Fast and easy programming of complex processes via AtmoCONTROL by means of the mouse or touchpad on the notebook.
Programme functions SingleDISPLAY and TwinDISPLAY:
  • Reading out, managing and organising the data logger.
  • Saving the log memory in various formats.
  • Online monitoring of up to 32 connected appliances.
  • Optical alarms when the alarm limits individually set at the ControlCOCKPIT are exceeded.
  • Automatic alarm to one or several e-mail addresses.
Standard equipment:
Controller:

PID microprocessor control with integrated auto-diagnostic system
Interior:st. steel, material 1.4301 (ASTM 304), with all-round deep-drawn ribs to integrate the large-area heating with ceramic-metal sheath
Internals:st. steel grids (sizes 30 and 55: 1 grid, sizes 75 to 750: 2 grids)
Max. loading per grid:sizes 30 to 260: 20 kg
sizes 450 and 750: 30 kg
Housing:textured st. steel, rear zinc-plated steel, intuitively operated SingleDISPLAY (IN / IF) or TwinDISPLAY (INplus / IFplus) colour graphical display with Multi-Touchscreen;
inner glass door, outer st. steel door, fully insulated st. steel door (from size 450 two leaves)
Fresh air:admixture of pre-heated fresh air by electronically adjustable air flap
Connection:mains cable with plug (German type)
Installation:4 feet; sizes 450 and 750 mounted on lockable castors
Technical Data:
Min. working temperature range:5 K (IN / INplus), 10 K (IF / IFplus) above ambient temperature to + 80 °C
Setting temperature range:+ 20 to + 80 °C
Setting accuracy:0.1 K
  • TwinDISPLAY: ControlCOCKPIT with 2 TFT colour graphical displays with touchscreen.
  • Available parameters on the ControlCOCKPIT: temperature, fan speed, exhaust air flap position, programme time, illumination.
  • 2 Pt100 sensors DIN class A in a 4-wire circuit for mutual monitoring, taking over functions in case of an error.
  • HeatBALANCE function for application specific adjustment of heat output distribution (balance) between the upper and lower heating groups in an adjustment range between - 50 and + 50 %.
  • ControlCOCKPIT with USB port for uploading programmes, reading out protocol logs, activating the User-ID function.
  • Displaying of already logged protocol data on the ControlCOCKPIT (max. 10000 values correspond to approx. 1 week).
  • AtmoCONTROL software on an USB stick for programming, managing and transferring programmes via Ethernet interface or USB port.
  • Intuitive programming and archiving of ramps and programme sequences.
  • Synchronous visualisation of the created programme sequence during programming.
  • Application-specific repeat functions (Loops) can be inserted within a temperature control programme in any place.
  • Simple creation of repeating weekly programmes.
  • Ethernet interface on the rear of the appliance for reading out the protocol log and for uploading and implementing programmes and for online logging.
  • Multiple overtemperature protection: electronic temperature monitoring TWW/TWB (protection class 3.1 or 2) and mechanical temperature limiter TB (protection class 1) acc. to. DIN 12880, AutoSAFETY automatically adjusts to the set value within a freely adjustable tolerance range. Setting individual MIN / MAX values for over / undertemperature alarm.
 TypeChamber volume
l
Inside dimension
(W x D x H)
mm
Outside dimens­ions
(W x D x H)
mm
Load
capacity,
max.
kg
PUOrder No.in Stock  
IN30plus32400 x 250 x 320585 x 434 x 7046015432111-Login / Registration 
IN55plus53400 x 330 x 400585 x 514 x 7848015432112-Login / Registration 
IN75plus74400 x 330 x 560585 x 514 x 94412015432113-Login / Registration
 IN110plus108560 x 400 x 480745 x 584 x 86417515432114-Login / Registration
 IN160plus161560 x 400 x 720745 x 584 x 110421015432115-Login / Registration
 IN260plus256640 x 500 x 800824 x 684 x 118330015432116-Login / Registration
 IN450plus4491040 x 600 x 7201224 x 784 x 124730015432117-Login / Registration
 IN750plus7491040 x 600 x 12001224 x 784 x 172630015432118-Login / Registration
IF30plus32400 x 250 x 320585 x 434 x 7046015432119-Login / Registration 
IF55plus53400 x 330 x 400585 x 514 x 7848015432120-Login / Registration
IF75plus74400 x 330 x 560585 x 514 x 94412015432121-Login / Registration
IF110plus108560 x 400 x 480745 x 584 x 86417515432122-Login / Registration
IF160plus161560 x 400 x 720745 x 584 x 110421015432123-Login / Registration
IF260plus256640 x 500 x 800824 x 684 x 118330015432124-Login / Registration
IF450plus4491040 x 600 x 7201224 x 784 x 124730015432125-Login / Registration
IF750plus7491040 x 600 x 12001224 x 784 x 172630015432126-Login / Registration
Inside dimensions: depth less 39 mm for fan.
Outside dimensions: depth without door handle, depth of handle 56 mm. Height with size 450 and 750 with castors.
Further options such as equipment with intertior lighting or different ports are available on request. Extra equipment has to be stated with the order directly.
Accessories:
Accessories for series U, I, S, ICH, ICP, IPP, IPS, HPP, HCP / Memmert